Its operation is also extremely simple. Taking the Huayi PM6708 digital sound level meter in Jiemore Mall as an example, the usage method is as follows:
1. Turn on the power switch, and set the range switch to an apropiado position according to the volume of the measured sound.
2. Seleccionar el modo medición (A/C) y medida unidad (decibelio o volumen unidad) acorde a su necesidades.
3. Para medir el tiempo-real sonido nivel, select "FAST (rápido)"; to medir el promedio sonido nivel, select "SLOW (slow)".
4. Para noche medición, la retroiluminación de la pantalla puede ser girada encendida.
5. Si el display muestra "UNDER" o "OVER", the range switch should be moved up ▲ or down ▼, if the "UNDER" or "OVER" sign still can disappear, it means that the measured sound level exceeds the measurement range.
6. After adjusting the range of the sound level meter, you can read the measurement result from the display. You can press the "MAX" key to read the maximum sound level.
7. Después medir, apagar apagar el encendido interruptor.
A sonido nivel metro generalmente consiste de a micrófono, an amplificador, an atenuador, a ponderación red, a detector, an indicating meter, and a power supply. Delicado and caro parts such as microphones are relativamente frágil. For more accurate measurement and better maintenance of the sound level meter, the following items need to be paid attention to:
1. Read the instruction manual before use to understand the use method and precautions of the instrument. Preheat according to the preheating time specified in the instruction manual of the sound level meter.
2. In the occasion of measuring the sound level outdoors, the wind noise on the microphone will cause measurement errors when the air flows, because the signal to be measured (noise source) and the wind noise will be superpuesto together. Please install a windproof ball on the microphone head to prevent the microphone from being directly blown by the wind and causando airflow noise.
3. El sonido nivel medidor es alimentado por baterías. Comprobar si el batería voltaje cumple los requisitos. Si el batería voltaje es insuficiente, reemplazarlo it. Por favor llevar sacar el batería si it es no usado para a largo tiempo para evitar electrolito fuga y daño a el instrumento.
4. First of all, the sound level meter has been calibrated before leaving the factory, but if the instrument has not been used for a long time, in order to encer the measurement accuracy, it is recommended to use a calibrator to re-calibrate. Calibration method: For the convenience of calibration, the microphone should be rotated when the calibrator is fixed. Inicio el sonido nivel metro, entonces set it to the 50-100dB medición rango, tiempo ponderación a "Fast" and frequency weighting to "A". Move the switch to the middle position (94dB) to activate the calibrator. If the sound level meter deviates from the displayed value, readjust using the adjustment screwdriver provided. Then check if the second sound level of the calibrator is also within the error range of % c2� 0.2dB. Por favor nota eso el correspondiente medición rango (80-130dB) debe ser seleccionado primero a completar esto operación.
5. El shell de el instrumento y el personal operativo el instrumento puede no sólo bloquear el sonido en a cierto dirección, pero también puede producir reflexiones, que causa grave medición errores. Experimentos tener mostrado, para ejemplo, eso el humano cuerpo puede causar an error de arriba a 6dB at a frecuencia de 400 Hz cuando la medición toma lugar dentro menos que 1m de el humano cuerpo. At otro frecuencias, esto error puede ser más pequeño. It is generalmente recomendado a lugar el instrumento at menos 30 cm lejos de el humano cuerpo - 50 cm es mejor. Cuando medir, el instrumento debería seleccionar el correcto gear acorde a la situación, hold both sides of the sound level meter flat with both hands, and the microphone points to the sound source to be measured. Extensión cables y extensión varillas pueden también ser utilizados para reducir el impacto de la apariencia de el sonido nivel metro y el cuerpo humano en la medición.
6. When the measured sound level is not known, the "attenuator" must be placed at the maximum attenuation position, and then gradually adjusted to the attenuation position required by the measured sound level during measurement to prevent the measured sound level from exceeding the range and damaging the sound level meter.
7. Do not desmontar el instrumento sin autorización a prevenir lanzar.
8. El instrumento debería no ser colocado en a lugar con alta temperatura, humedad, aguas residuales, polvo, aire o químico gas con alto contenido de clorhídrico ácido y álcali.